| Operation Range | 0 to 207 kPa (0 to 30 psia); approximately 0 to 9 m (0 to 30 ft) of water depth at sea level, or 0 to 12 m (0 to 40 ft) of water at 3,000 m (10,000 ft) of altitude |
| Factory Calibrated Range | 69 to 207 kPa, 0° to 40°C 10 to 30 psia, 32° to 104°F |
| Burst Pressure | 310 kPa or 18 m depth 45 psia or 60 ft depth |
| Water Level Accuracy* | Typical error: ±0.1% FS, 1.0 cm (0.03 ft) Water Maximum error: ±0.2% FS, 2.0 cm (0.06 ft) Water |
| Raw Pressure Accuracy** | ±0.3% FS, 0.62 kPa (0.09 psi) Maximum error |
| Resolution | <0.02 kPa Water 0.007 ft Water |
| Pressure Response Time (90%)*** | <1 second at a stable temperature; measurement accuracy also depends on temperature response time |
| Operation Range | 0 to 400 kPa (0 to 58 psia); approximately 0 to 30.6 m (0 to 100 ft) of water depth at sea level, or 0 to 33.6 m (0 to 111 ft) of water at 3,000 m (10,000 ft) of altitude |
| Factory Calibrated Range | 69 to 400 kPa (10 to 58 psia), 0° to 40°C (32° to 104°F) |
| Burst Pressure | 500 kPa (72.5 psia) or 40.8 m (134 ft) Depth |
| Water Level Accuracy* | Typical error: ±0.1% FS, 3.0 cm (0.1 ft) Water Maximum error: ±0.2% FS, 6.0 cm (0.2 ft) Water |
| Raw Pressure Accuracy** | ±0.3% FS, 1.20 kPa (0.17 psi) Maximum error |
| Resolution | <0.04 kPa (0.006 psi), 0.41 cm (0.013 ft) Water |
| Pressure Response Time (90%)*** | <1 second at a stable temperature; measurement accuracy also depends on temperature response time |
| Operation Range | 0 to 145 kPa (0 to 21 psia); approximately 0 to 4 m (0 to 13 ft) of water depth at sea level, or 0 to 7 m (0 to 23 ft) of water at 3,000 m (10,000 ft) of altitude |
| Factory Calibrated Range | 69 to 145 kPa (10 to 21 psia), 0° to 40°C (32° to 104°F) |
| Burst Pressure | 310 kPa (45 psia) or 18 m (60 ft) Depth |
| Water Level Accuracy* | Typical error: ±0.1% FS, 0.4 cm (0.013 ft) Water Maximum error: ±0.2% FS, 0.8 cm (0.026 ft) Water |
| Raw Pressure Accuracy** | ±0.3% FS, 0.43 kPa (0.063 psi) Maximum error |
| Resolution | <0.014 kPa (0.002 psi), 0.14 cm (0.005 ft) Water |
| Pressure Response Time (90%)*** | <1 second at a stable temperature; measurement accuracy also depends on temperature response time |
| Operation Range | -20° to 50°C -4° to 122°F |
| Accuracy | ±0.2°C from 0° to 50°C ±0.36°F from 32° to 122°F |
| Resolution | 0.10°C at 25°C 0.18°F at 77°F |
| Response Time (90%) | 10 minutes in water (typical) |
| Stability (Drift) | 0.1°C per year 0.18°F per year |
| Wireless Standard | Bluetooth 5 LE |
| Transmission Range | Approximately 30.5 m (100 ft) line-of-sight |
| Radio Power | 1mW (0 dBm) |
| Real-time Clock | ±1 minute per month 0° to 50°C (32° to 122°F) |
| Battery | 2/3 AA, 3.6 Volt lithium, factory-replaceable |
| Battery Life (Typical Use) | 3 years with 1-minute logging and Bluetooth always on 7 years with 1-minute logging and Bluetooth off and LED off 10 years with 5-minute logging and Bluetooth off and LED off |
| Battery Indication | Battery status shown in HOBOconnect app (approximate) |
| Memory (Non-volatile) | Approximately 400,000 sets of pressure and temperature measurements |
| Data Download Time (with Bluetooth 5+ devices) | Approximately 3.5 minutes; 100,000 measurements < 1 minute |
| Weight | Approximately 154 g (5.43 oz) in air Approximately 53.9 g (1.9 oz) in fresh water |
| Dimensions | 3.18 cm (1.25 inches) diameter, 15.24 cm (6.0 inches) length; mounting hole 6.3 mm (0.25 inches) diameter |
| Wetted Materials | Polypropylene housing and lanyard; Viton and Buna-N O-rings; ceramic sensor in acetal end cap; stainless steel screws suitable for saltwater |
| Logging Modes | Fixed-interval (normal or statistics), burst or multiple intervals with up to 8 user-defined intervals and durations; logging intervals from 1 second to 18 hours. Refer to the HOBOconnect User’s Guide for details. |
| Start Modes | Immediate, next interval, date & time, button start (fob or app) |
| Stop Modes | Memory full, never stop (overwrite oldest data), date & time, or after set logging period |
| Pause | Pause and restart with HOBOconnect app |
| Alarms | Log high- or low-alarm events; alarm status shown in HOBOconnect app |
| Environmental Rating | IP68 |