Online Ultrasonic Thickness Gauge TIME®2131

 

Features

● Auto recognition of probe

● Dynamic compensation of measurement error caused by probe change and coupling condition

● Real time display coupling condition

● OLED screen with high contrast and brightness suitable to use in sunlight

● Scanning measuring mode, up to 20 times per second

● I-E testing mode, E-E testing mode and auto mode for option

● Memory of 3000 data: stored in 30 thickness files, 100 thickness values for each file

● Upgrade online, upgrade the firmware of the unit by WIFI or APP

● Dataview available for selection of data store, ZW5P, TSTU32 etc.

Spesifikasi

Measuring range (steel)

5PΦ10, 5PΦ10/90: 1.2-225.0 mm,
7PΦ6: 0.75-60.0 mm, ZW5P: 4.0-80.0
mm, TSTU32: 3.0-300.0 mm

Accuracy (steel)

ZW5P, TSTU32: ±0.10 mm (H<10.00
mm)±(1%H+0.01) mm (H≥10.0 mm)
5PΦ10, 5PΦ10/90, 7PΦ6: ±0.05 mm
(H<10.00 mm) ±(0.5%H+0.01) mm
(H≥10.00 mm)

Repeatability (steel)

ZW5P, TSTU32: 0.10 mm, 5PΦ10,
5PΦ10/90, 7PΦ6: 0.03 mm

Stability (steel)

ZW5P, TSTU32: 0.10 mm, 5PΦ10,
5PΦ10/90, 7PΦ6: 0.05 mm

Accuracy for wall thickness of curve surface(steel)

±0.1 mm

Sound velocity range (m/s)

508-18699 (m/s)

Accuracy for thickness when change sound velocity

≤±0.5 mm

Resolution

0.1 mm, 0.01 mm, 0.001 mm

Special display

Min. value, Max. Value, Average value

Alarm

Upper, lower limit

Two point calibration

mm/inch

Indicator for battery
power

Indicator for coupling
condition

Memory

6 sound velocity value, 3000 thickness
value

Communication

USB or WIFI

Switch off

Auto switch off in 2 minutes if no use

Working environment

Temperature: 0-40℃, Humidity: 90%
RH

Memory environment

Temperature: -25-60℃, Humidity: 90%
RH

Power

2 pcs AA battery

Working electricity

60 mA (3.0 V)

 